Measuring by comparison
Feeler gaugeFeeler gauge
Also called: thickness gauge (leaf type), clearance gauge, feeler gauge

Standards, measuring range and accuracy
- Standard
- JIS B 7524:2008 Feeler gauges[1]One source / reference
- Measuring range
- By manufacturer
- Scope of JIS B 7524 Thickness up to 3 mm, length up to 300 mm
- Niigata Seiki (combination sets) Leaf thickness 0.01–3.00 mm, leaf length 75–300 mm
- Shinwa Measuring D (25-leaf set) Leaf thickness 0.03–1.00 mm, leaf length 65 mm
- By manufacturer
- Resolution / graduation
- Description No graduations. The increments you can measure depend on the combination of leaves you have (example, 25-leaf set: 0.01 mm steps from 0.03–0.15 mm, 0.05 mm steps from 0.20–0.50 mm, 0.1 mm steps from 0.60–1.00 mm)
- Accuracy (instrument error / maximum permissible error)
- By manufacturer
- Shinwa Measuring Thickness Gauge (single leaf) Thickness 0.03–0.15 mm: ±0.005 mm; over 0.15 up to 0.50 mm: ±0.012 mm; over 0.50 up to 3.00 mm: ±0.020 mm
- Niigata Seiki graduated leaf (0.1 mm thick) Thickness ±0.01 mm
- By manufacturer
Metric ⇄ Inch & ASME/ISO Quick Reference: 1 mm = 0.03937 in (25.4 mm = 1 in) · 0.01 mm ≈ 0.0004 in (“4 tenths”) · 0.001 mm (1 µm) ≈ 0.00004 in (“40 millionths”). Standard reference temperature is 20 °C (68 °F) per ISO 1 / ASME B89.6.2. Related US standards: ASME B89.1.14 (calipers), ASME B89.1.13 (micrometers), ASME B89.1.10M (dial indicators), ASME B89.1.9 (gage blocks), ASME B89.3.7 (granite surface plates), ASME B1.2 (UN thread gages).
Types
Combination type (leaves bundled with a screw)
Most bundle about 9–25 leaves. Leaf length is 65–300 mm and width is 12.7–13 mm. Removing the screw lets you take the leaves off one at a time[2][3]
Single leaves and graduated leaves
Leaves are also sold one at a time, so you can replace just the damaged one. Some leaves have a 1 mm scale that shows how far they are inserted[4][5]One source / reference
Roll-type feeler gauge
Cut a long strip to the length you need. For assembly and adjustment work[6]One source / reference
Part names and functions
How to read the scale
- Read the nominal size (thickness) marked on the leaf. When two or more leaves are stacked, the sum of their nominal sizes is the gap sizeExample: if 0.30 mm and 0.05 mm leaves stacked together fit, the gap is 0.35 mm (calculated from the rules in the source)[3]
- The gap lies between the leaf that fits and the leaf that does notExample: if 0.10 mm fits and 0.11 mm does not, the gap is 0.10 mm or more and less than 0.11 mm (calculated from the rules in the source)[3][2]
Usage steps
- Clean the leaves and the area to be measuredClean the leaves well and apply a thin film of oil before use. Wipe off dirt with a soft cloth; do not use volatile solvents such as thinner, or alcohol. The white powder on new carbon tool steel leaves is a rust inhibitor[3][2]
- Insert a leaf into the gap, and take the thickness of the thickest leaf (or combination) that goes in with slight resistance as the gap size. Niigata Seiki has published a video, "How to use feeler gauges"[8]
- Do not force the leaf inForcing it in will get even a leaf thicker than the gap to go in. Judge by the thickness that goes in with light resistance[3]
- When stacking, use as few leaves as possible and sandwich thin leaves between thick onesTolerance is set per leaf, and stacking adds the tolerances up[3]
- Use together with a square or straightedgeOne use is to check the gap that appears when a square is placed against a workpiece, using a feeler gauge[9]
Printed from Kezuriba (kezuriba.net/en/gauges/feeler-gauge/)
Common measurement mistakes and countermeasures
Forcing the leaf in and calling it "fits"
WhyIf you push, even a leaf thicker than the gap can seem to fit. It can also damage the leaf
CountermeasureJudge by the thickness that goes in with light resistance. Do not force it in[3]
Stacking many leaves to measure
WhyTolerance is a per-leaf value, and stacking adds the tolerances up. Three ±0.012 mm leaves can be off by up to ±0.036 mm (calculated from Shinwa Measuring's tolerance)
CountermeasureUse as few leaves as possible. Sandwich thin leaves between thick ones[3]
Care, storage, inspection and calibration
Care and storage
- After use, clean it and store it with a thin film of oil. Carbon tool steel leaves rust easily[3][2]
- Store away from direct sunlight and from hot or humid places[3]
- Do not subject it to strong shocks such as dropping[3]
- If you use it where rust is likely, stainless steel leaves are an option[7]
Inspection and calibration
- Tolerance is set per product (Shinwa Measuring leaves are ±0.005 to ±0.020 mm depending on thickness). Before use, check that it is accurate enough for the gap you want to measure[3]
- Bent, scratched or rusty leaves can be replaced one at a time by removing the screw (single leaves are sold)[3][4]
Beware of name confusion: Mitutoyo's 'thickness gauge' (shikkunesu gauge) is a dial-type thickness gauge that clamps between an anvil and a spindle (Mitutoyo has a usage page), and is a different thing from the JIS 'feeler gauge' (leaf type). Niigata Seiki's 'thickness gauge' (shikunesu gauge) is used as a name for the leaf type.
Related instruments
Frequently asked questions about Feeler gauge
- What should I check first before measuring with Feeler gauge?
- Clean the leaves and the area to be measured Clean the leaves well and apply a thin film of oil before use. Wipe off dirt with a soft cloth; do not use volatile solvents such as thinner, or alcohol. The white powder on new carbon tool steel leaves is a rust inhibitor Always wipe dust, burrs, and cutting oil from both the measuring faces and the workpiece before taking a reading.
- What is the most common mistake when using Feeler gauge, and how do I prevent it?
- Common error: Forcing the leaf in and calling it "fits". If you push, even a leaf thicker than the gap can seem to fit. It can also damage the leaf Countermeasure: Judge by the thickness that goes in with light resistance. Do not force it in
- How should Feeler gauge be cared for, stored, and inspected?
- Care and storage: After use, clean it and store it with a thin film of oil. Carbon tool steel leaves rust easily Inspection and calibration: Tolerance is set per product (Shinwa Measuring leaves are ±0.005 to ±0.020 mm depending on thickness). Before use, check that it is accurate enough for the gap you want to measure
Sources
Content without a mark was confirmed by two or more sources from different publishers. Procedures and cautions are summarized in Kezuriba's own words based on the sources, and the figures were drawn by Kezuriba. Accuracy values are manufacturers' published figures unless noted otherwise (values taken from JIS tables are marked as such).
- Japanese Standards Association (JSA), "JIS B 7524:2008 Feeler gauges (bibliographic record)"
- Niigata Seiki, "Thickness Gauge (Product Information)"
- Shinwa Measuring, "Thickness Gauge D 65mm 25-Leaf Set (Part No. 73782) Instruction Manual"
- Niigata Seiki, "Thickness Gauge Leaf 75mm (Product Information)"
- Niigata Seiki, "Graduated Thickness Gauge Leaf 100mm (Product Information)"
- Niigata Seiki, "Feeler Gauge FG Series (Product Information)"
- Niigata Seiki, "Stainless Steel Thickness Gauge (Product Information)"
- Niigata Seiki, "Measuring Tools How-to Videos (list)"
- Obishi Keiki, "I-Type Square Instruction Manual FG202"