Measuring length

Outside micrometerOutside micrometer

Also called: micrometer, mic, standard outside micrometer, Digimatic micrometer

Illustration of using an outside micrometer

Standards, measuring range and accuracy

Standard
  • JIS B 7502 Geometrical product specifications (GPS) – Dimensional measuring equipment – Micrometers (the current JSA edition is 2026; Mitutoyo's explanation covers the 2016 edition)
  • ISO 3611:2023 Design and metrological characteristics of micrometers for external measurements
  • ASME B89.1.13 Micrometers
[1][2][3][4]
Measuring range
  • General The spindle travel is 25 mm. Ranges go in 25 mm steps: 0–25, 25–50, 50–75 mm ..., up to 0–2000 mm
  • Selection examples To measure around 95 mm, use the 75–100 mm micrometer
  • By manufacturer
    • Mitutoyo 103-137/103-138 0–25 / 25–50 mm
    • Teclock SSM-750/850 0–30 / 30–66 mm (digital, 10 mm per revolution)
[6][7][9][5][12]
Resolution / graduation
  • Standard Graduation 0.01 mm (a scale line is about 2 µm wide in terms of reading, about 1/5 of the interval, so you can estimate by eye down to 0.001 mm)
  • With vernier 0.001 mm
  • Digital 0.001 mm (some models 0.0001 mm)
  • Screw thread The standard screw has a 0.5 mm pitch. One thimble revolution advances the spindle 0.5 mm. Some models have a 2 mm pitch for rapid feed
[3][6][7][10]
Accuracy (instrument error / maximum permissible error)
  • Approach
    • In JIS B 7502 (2016), "instrument error" became "indication error", and the most important indication error is the "full surface contact error J". The upper limit is set by the maximum permissible error J_MPE
    • J is found by setting the zero point at the minimum measuring length using the constant-pressure device, then clamping a JIS B 7506 grade 0 or grade 1 gauge block across the entire measuring faces and subtracting the gauge dimension from the indicated value
    • The main factors that determine accuracy are screw accuracy, flatness of the measuring faces, parallelism of the measuring faces, and frame rigidity
  • By manufacturer
    • Mitutoyo 103-137 (0–25 mm, 0.01 mm) J_MPE ±2 µm, parallelism 2 µm, flatness 0.6 µm, measuring force 5–10 N
    • Teclock SSM-750 (0–30 mm, 0.001 mm digital) Indication error ±2 µm
[3][9][5][12][1]

Metric ⇄ Inch & ASME/ISO Quick Reference: 1 mm = 0.03937 in (25.4 mm = 1 in) · 0.01 mm ≈ 0.0004 in (“4 tenths”) · 0.001 mm (1 µm) ≈ 0.00004 in (“40 millionths”). Standard reference temperature is 20 °C (68 °F) per ISO 1 / ASME B89.6.2. Related US standards: ASME B89.1.14 (calipers), ASME B89.1.13 (micrometers), ASME B89.1.10M (dial indicators), ASME B89.1.9 (gage blocks), ASME B89.3.7 (granite surface plates), ASME B1.2 (UN thread gages).

Types

Standard outside micrometer (analog, graduation 0.01 mm)

[5][6][7]

With vernier (graduation 0.001 mm)

A vernier scale above the reference line on the sleeve lets you read 0.001 mm directly[3][8]One source / reference

With mechanical counter (digit type)

[3][9]One source / reference

Digital (Digimatic, resolution 0.001 mm; high-accuracy 0.0001 mm models also exist)

[10][9][11][12]

Constant-pressure device types: ratchet stop / friction thimble (F-type) / ratchet thimble

The ratchet stop can be recognized by sound and feel, but it gives small clicking shocks and is not suited to one-handed operation. The friction thimble gives neither sound nor shock and can be used one-handed. The ratchet thimble can be used one-handed and confirmed by sound[3][9][7][6]

Special micrometers: point, blade, spline, single-ball and ball-type (pipe wall), U-shaped sheet-steel, disc (gear tooth thickness), thread (interchangeable anvils), V-anvil, ball-type tooth thickness, caliper-type inside, etc.

[3][9][7][11]

Interchangeable-anvil type (wide range)

[9]One source / reference

Part names and functions

#NameFunction
1FrameThe U-shaped body. It holds the anvil and spindle. Its rigidity affects accuracy.[13][6][11]
2AnvilThe fixed measuring face. Often carbide-tipped.[13][6][11]
3SpindleThe movable measuring face, advanced and retracted by a screw. It is on the same axis as the anvil.[13][6][11]
4Measuring facesFlatness and parallelism determine accuracy. Carbide is common; hardened steel if cost is the priority.[13][9]
5Clamp (spindle clamp)Locks the spindle. Loosen it for storage.[13][6]
6Sleeve (barrel), reference line, and sleeve scaleA 1 mm scale above the reference line and a 0.5 mm scale below it.[13][6][11]
7Thimble and thimble scaleA scale dividing the circumference into 50 parts. One division is 0.01 mm.[13][6][11]
8Ratchet stop (constant-pressure device)Slips once a set measuring force is exceeded, evening out the force.[13][6][11]
9Heat shield (insulating cover)Reduces heat transfer from the hand to the frame. Hold the micrometer here.[13][9][11]
10Speeder (rapid-feed knob)A part for advancing the spindle quickly.[9][3]
11ZERO/ABS, ORIGIN, HOLD buttons, display (digital types)Origin setting, relative zero, and hold.[13][9][11]

How to read the scale

  1. On the sleeve, read the largest graduation visible to the left of the thimble edge (check both the 1 mm scale above and the 0.5 mm scale below)Example: the 7 mm line is visible but the 7.5 mm line is not → 7.0 mm[6][7][3]
  2. On the thimble scale, read the line that aligns with the sleeve reference line. If it is between lines, read the line below the reference lineExample: thimble 37 aligns with the index line → 0.37 mm[7][6][3]
  3. Add the twoExample: 7.0 + 0.37 = 7.37 mm (Mitutoyo). Niigata University's example reads to half a division by eye: 7.000 + 0.375 = 7.375 mm, and if the 0.5 mm line is visible, 7.5 + 0.375 = 7.875 mm[3][6]
  4. With vernier (0.001 mm): add in the order sleeve → thimble → vernier coincident lineExample: sleeve 6 mm + thimble 0.21 mm (index line between 21 and 22) + vernier 0.003 mm = 6.213 mm[3]
  5. Mechanical counter type: add the vernier's 0.001 mm to the 4 digits of the counterExample: 2.99 mm (counter) + 0.004 mm (vernier) = 2.994 mm[3]
  6. Rule of thumb for reading down to 1 µm by eye: a scale line is about 2 µm wide in terms of reading (about 1/5 of the 10 µm interval). Judge 1 µm or 2 µm from how the reference line overlaps the thimble line[13][3][14]
  7. Read the scale from the front, directly above the point where the reference line and the thimble line coincide[7][13][3]
How to read an outside micrometer (sleeve + thimble)
Sleeve (upper = 1 mm, lower = 0.5 mm)05Thimble (1 division = 0.01 mm)3540← aligned with the index line

Sleeve 7.0 mm + thimble 0.37 mm (37 divisions) = 7.37 mm

On the sleeve, the upper side has 1 mm graduations and the lower side has 0.5 mm graduations. The thimble divides 0.5 mm per revolution into 50 parts (1 division = 0.01 mm).

Usage steps

  1. Selection: choose by checking type, measuring range, and accuracy[3][7]
  2. Temperature soaking: let the micrometer and workpiece stabilize thoroughly at room temperature[7][3]
  3. Cleaning: clamp a sheet of white paper (lint-free) between the anvil and spindle measuring faces and pull it out gently to remove dirt. Also wipe chips off the spindle's outer surface[7][13][3][6]
  4. Operation check: turn the thimble through its full range and check for sticking or uneven movement[13]
  5. Zero-point check: turn the ratchet stop slowly to bring the measuring faces together and check that the thimble zero line aligns with the sleeve reference lineMitutoyo says to confirm with 1.5–2 turns of the ratchet and 3–5 applications of constant pressure (for 0–25 mm models). For models above 25 mm, use a micrometer standard rod. If it is off, turn the sleeve with a hook spanner to match, or use the offset as a correction value[7][13][6]
  6. When mounting in a stand, clamp the center of the frame and do not overtighten. Fix it at an angle where the scale can be seen evenly from above and below[7][3]
  7. Measurement: in the same posture and conditions as the zero-point check, lightly touch both measuring faces to the workpiece, turn the ratchet stop slowly 3–5 times, and readWhen approaching the workpiece, always turn the ratchet stop. Turning the thimble directly applies excessive measuring force and wears the screw, which degrades accuracy[7][6]
  8. Reading: read from the front. If you used the clamp, release it after reading[7][13]
  9. For long sessions, check the zero point periodically to allow for thermal expansion[13][14]

How errors come about

The "Common measurement mistakes" below cover what you end up doing wrong, while this section covers why the reading drifts and by how much.

1Follows the Abbe principle: tilt error becomes "second-order"

  • The spindle (the screw that carries the scale) and the line of measurement are on one straight line, so it follows the Abbe principle. Even with a slight tilt the error is small, L(1 − cos θ) ≈ L·θ²/2, a "second-order error" proportional to the square of the tilt. This is an order of magnitude different from a caliper's "first-order error" h·θ, which is directly proportional to the tilt.
  • That is why a resolution of 0.001 mm is meaningful. However, if you tilt an inside micrometer in a hole, it reads larger by √(ℓ² + X²) − ℓ (with ℓ = 200 mm and one end off by 10 mm, about 0.25 mm; calculated example).

Worked ExampleWith a tilt of 0.01 mm per 50 mm (θ = 0.0002 rad), the error at L = 25 mm is about 0.0000005 mm (0.5 nm). Compared with the caliper-tip value of 0.008 mm, it is almost negligible (calculated example)

Abbe principle: on a caliper the scale and the measuring point are apart; on a micrometer they are in one line
Caliper: the scale and the measuring point are h apartIf the jaw tilts by θ, error f = h × θ (first-order error)At a tilt of 0.01 mm/50 mm: h 40 mm → 8 µm, h 10 mm → 2 µmMicrometer: the scale and the line of measurement are in one lineError ≈ L × θ²/2 (second-order error)At the same tilt, L 25 mm → about 0.5 nm (calculated example)Graduation lines of the main scaleMeasuring point (jaw tip)hAxis of the scale (screw) = line of measurement (follows the Abbe principle)Clamp the workpiece at the root of the jaws (to keep h small). For shapes reachable only with the tips, switch to a micrometer or a dedicated measuring instrumentValues are from Mitutoyo's example (h 40 mm) and values calculated with the same formula. The picture is a conceptual illustration generated in Canva

[3][18]Paper

2Measuring force: even out the "way you apply it" with the ratchet

  • The constant-pressure device (ratchet stop, friction thimble, ratchet thimble) slips once a set force is exceeded and evens out the measuring force. For example, the measuring force of the Mitutoyo 103-137 is 5–10 N.
  • An engineering-education study comparing the operation of skilled people and beginners concluded that keeping the measuring force constant when bringing the spindle against the part matters more than reading the scale. Once you have learned to read it, the next step is practicing applying it evenly.
  • When a ball or cylinder is clamped between flat measuring faces, the contact areas are squashed by the measuring force and it reads smaller than the actual size (Hertz formula). A φ10 mm steel ball clamped at 5 N reads about 1.1 µm smaller, and at 10 N about 1.8 µm smaller (calculated example).

Worked ExampleSteel ball: δ₁ = 0.82 ∛(P²/D) µm; steel cylinder: δ₂ = 0.094 (P/L) ∛(1/D) µm (P: measuring force in N, D: diameter in mm, L: cylinder length in mm; Mitutoyo's formulas)

Calculate the squashing from measuring force (steel)

Apparent reduction δ = 1.11 µm

Measuring force and squashing: clamping a ball between flat measuring faces makes it read smaller by δ
Spindle measuring faceMeasuring face of the anvilDSquashing (δ in total, top and bottom)δ₁ = 0.82 ∛(P²/D) µmP: measuring force (N), D: ball diameter (mm), steelφ10 mm, 5 N → about 1.1 µmφ10 mm, 10 N → about 1.8 µmThe squashing is drawn exaggerated

An example calculated from the Hertz formula in Mitutoyo's material (steel, Young's modulus 205 GPa). For a cylinder, δ₂ = 0.094 (P/L) ∛(1/D) µm (L: cylinder length in mm). Even with a measuring force of 5–10 N like a micrometer's, measuring a ball reads smaller by about 1 µm.

[3][19][20]PaperReports from public institutions

3Temperature: heat from your hands, and workpieces fresh from cutting

  • Holding the frame bare-handed makes it expand from the heat of your hand and shifts the zero point. In Mitutoyo's experimental graph, the 300 mm size expanded about 14 µm in 30 minutes and even the 50 mm size about 4 µm (experimental values, not guaranteed values). Hold the heat shield, or use it mounted on a stand.
  • In one example, a 200 mm standard rod held in a hand at 31 °C lengthened about 18 µm in 10 minutes (same source). Wear gloves when holding a standard rod, or support it lightly by its insulated part.
  • Measuring a 35 °C steel workpiece (100 mm) fresh from cutting with a 20 °C micrometer reads about 0.018 mm larger (calculated example). Half of the 35 µm tolerance of a φ100 h7 is wiped out by temperature alone.
A micrometer expands from hand heat: gripping it bare-handed or holding the heat shield
× Gripping the frame bare-handedThe 300 mm size expands about 14 µm in 30 minutesAbout 4 µm even on the 50 mm size (Mitutoyo's experimental values)○ Holding the heat shieldHeat transfers less, so the zero point moves lessFor long measuring sessions, mount it on a standValues read from the experimental graph, not guaranteed values. The picture is a conceptual illustration generated in Canva
Open these conditions in the thermal expansion calculator →

[3][21]Patent

4Support and sag (large micrometers)

  • On a micrometer with a large measuring range, the frame sags depending on how it is supported and its orientation, moving the zero point by several µm to over ten µm. In Mitutoyo's table, changing the support on a 925 mm maximum-measuring-length size gives −22.5 µm in one example.
  • Set the zero point in the same posture and support as when actually measuring.

[3]

Article "Measurement error basics", which covers the Abbe principle, 20 °C, measuring force, uncertainty, and pass/fail decisions →

Common measurement mistakes and countermeasures

Heat from the hand

WhyHolding the frame bare-handed makes it expand from the heat and shifts the zero point. In Mitutoyo's experimental graph it is about 14 µm in 30 minutes for the 300 mm size and about 4 µm even for the 50 mm size. The standard rod also lengthens from hand temperature (holding a 200 mm rod for 10 minutes with a hand at 31 °C gives about 18 µm)

CountermeasureHold the heat shield. Hold the standard rod with gloves or support it lightly by its insulated part. A lengthened rod takes time to return to its original length[3][7]

Temperature difference between the micrometer and the standard rod

WhyIf they are kept in different places, their temperatures differ and the zero-point setting is off

CountermeasureKeep both in the same place for at least several hours before setting the zero point[3]

Excessive measuring force / turning the thimble directly

WhyIt damages the screw and also deforms the workpiece. Turning it roughly presses too hard on the anvil side

CountermeasureUse the constant-pressure device correctly. Make contact slowly[6][7][3]

Parallax

WhyThe coincident position of the scales appears to change with eye position

CountermeasureRead directly above the point where the reference line and the thimble line coincide[7][3][13]

Missing the 0.5 mm

WhyIf you overlook whether the 0.5 mm line below the sleeve is visible, you are off by 0.5 mm

CountermeasureCheck both the upper and lower scales (a micrometer head with 1 mm pitch avoids this misreading)[6][3]

Posture and support of large micrometers

WhyThe frame sags depending on how it is supported and its posture, shifting the zero point by several µm to over ten µm (in Mitutoyo's table, changing the support on a 925 mm size gives −22.5 µm in one example)

CountermeasureSet the zero point in the same posture and support as the actual measurement[13][3]

Workpiece flattening with point or line contact

WhyClamping a ball or cylinder between flat faces deforms it elastically under the measuring force, so it reads smaller than it is (Hertz formula)

CountermeasureIf needed, estimate the deformation (for a steel ball, δ = 0.82 ∛(P²/D) µm; about 1.1 µm for φ10 mm at 5 N). Choose an instrument with a small measuring force[3]

Care, storage, inspection and calibration

Care and storage

Inspection and calibration

A micrometer follows the Abbe principle, placing the workpiece and the standard scale (the screw) on one line in the measuring direction, so it measures more accurately than a caliper (Mitutoyo). The three-wire method for the effective diameter of a thread (E = M − 3d + 0.866025P, best wire diameter 0.577P) is also done with a micrometer.

Related instruments

Frequently asked questions about Outside micrometer

What should I check first before measuring with Outside micrometer?
Selection: choose by checking type, measuring range, and accuracy Always wipe dust, burrs, and cutting oil from both the measuring faces and the workpiece before taking a reading.
What is the most common mistake when using Outside micrometer, and how do I prevent it?
Common error: Heat from the hand. Holding the frame bare-handed makes it expand from the heat and shifts the zero point. In Mitutoyo's experimental graph it is about 14 µm in 30 minutes for the 300 mm size and about 4 µm even for the 50 mm size. The standard rod also lengthens from hand temperature (holding a 200 mm rod for 10 minutes with a hand at 31 °C gives about 18 µm) Countermeasure: Hold the heat shield. Hold the standard rod with gloves or support it lightly by its insulated part. A lengthened rod takes time to return to its original length
How should Outside micrometer be cared for, stored, and inspected?
Care and storage: After use, check for damage and clean; if water-soluble cutting fluid got on it, apply rust prevention Inspection and calibration: Flatness of the measuring faces: place an optical flat on the anvil and count the red interference fringes under white light. One fringe ≈ 0.32 µm (e.g., 4 fringes is about 1.3 µm)

Sources

Content without a mark was confirmed by two or more sources from different publishers. Procedures and cautions are summarized in Kezuriba's own words based on the sources, and the figures were drawn by Kezuriba. Accuracy values are manufacturers' published figures unless noted otherwise (values taken from JIS tables are marked as such).

  1. Japanese Standards Association (JSA), "JIS B 7502:2026 Geometrical product specifications (GPS) – Dimensional measuring instruments – Micrometers (bibliographic record)"
  2. ISO, "ISO 3611:2023 Design and metrological characteristics of micrometers for external measurements"
  3. Mitutoyo, "Quick Guide to Precision Measuring Instruments (Catalog No. E11003(7), as of August 2024)"
  4. ASME, "ASME B89.1.13 Micrometers"
  5. Mitutoyo, "Standard Outside Micrometer M110 103-137 Specifications"
  6. Niigata University, "Niigata University Faculty of Engineering, Technical Staff Training Text, Chapter 1: Workshop Measurement"
  7. Mitutoyo, "How to Use Micrometers Correctly and Cautions"
  8. Mitutoyo, "Outside Micrometer M310/M320 102-301 Specifications"
  9. Mitutoyo, "Mitutoyo Product Fundamentals (Bulletin No. 2180, Mitutoyo America)"
  10. Mitutoyo, "What Is Measurement (sokutei)? The Difference Between Sokutei and Keisoku"
  11. Niigata Seiki, "Digital Micrometer QEM29 Series Instruction Manual"
  12. Teclock, "General Catalog 5: Micrometers"
  13. Mitutoyo, "Check Points for Measuring Instruments (Catalog No. E12024)"
  14. Mitutoyo, "How to Use Interchangeable-Rod Depth Micrometers"
  15. Japanese Standards Association (JSA), "JIS B 0641-5:2021 Uncertainty in the conformity testing of indicating measuring instruments, preview"
  16. Gunma Industrial Technology Center, "Kawahara, Masuda, 'Uncertainty evaluation for ISO/IEC 17025 and JCSS registration renewal,' research report (2016)"
  17. Gunma Industrial Technology Center, "Oyake, Masuda, Nakamura, 'Establishing on-site calibration methods for calipers and micrometers,' research report (2023)"
  18. P. Schellekens et al. (Annals of the CIRP 47(2), 1998), "Design for Precision: Current Status and Trends"
  19. Takeo and Xia (Journal of Japanese Society for Engineering Education 61(6), 2013), "A study on educational practice of keeping the measuring force of outside micrometers constant"
  20. National Institute of Standards and Technology (NIST), "T. Doiron and J. Beers, 'The Gauge Block Handbook' (NIST Monograph 180)"
  21. Mitutoyo Corporation, "US 8,245,413 B2, "Heat insulating cover and micrometer""