Angles and flatness

Precision levelPrecision level (spirit level)

Also called: block level, flat-base level, level

Illustration of using a precision level

Standards, measuring range and accuracy

Standard
JIS B 7510:1993 Precision levels[1][2]
Resolution / graduation
  • Sensitivity 0.02 / 0.05 / 0.1 mm/m (grade A square type)
  • Definition The inclination needed to move the bubble one division. Expressed as height per 1 m of base. 1 arcsecond = 4.85 µm per m (≈ 5 µm). 0.02 mm/m ≈ 4 arcseconds, 0.05 mm/m ≈ 10 arcseconds, 0.1 mm/m ≈ 20 arcseconds
[2][3]One source / reference

Metric ⇄ Inch & ASME/ISO Quick Reference: 1 mm = 0.03937 in (25.4 mm = 1 in) · 0.01 mm ≈ 0.0004 in (“4 tenths”) · 0.001 mm (1 µm) ≈ 0.00004 in (“40 millionths”). Standard reference temperature is 20 °C (68 °F) per ISO 1 / ASME B89.6.2. Related US standards: ASME B89.1.14 (calipers), ASME B89.1.13 (micrometers), ASME B89.1.10M (dial indicators), ASME B89.1.9 (gage blocks), ASME B89.3.7 (granite surface plates), ASME B1.2 (UN thread gages).

Types

Square type (faces ensured parallel and square, with V-groove) / flat type

[2][3]One source / reference

Part names and functions

#NameFunction
1Main vial (bubble length adjustable with the air chamber)Reads tilt from the bubble position. The bubble length can be adjusted in the bubble chamber[2]
2Secondary vialChecks level in the roll direction when placed on a cylinder.[2]
3Adjusting screwAdjusts the bubble position (zero point).[2]
4V-groove / base / gripThe base sits on a flat surface and the V-groove on a round bar or shaft. The handle is the part you hold[2]

How to read the scale

  1. Read the divisions at both ends of the bubble and take the average[2][4]
  2. Get the height difference from the sensitivityExample: nominal 200 (length 200 mm), sensitivity 0.02 mm/m → 1 division = 4 µm/200 mm[4]

Bubble Division to Tilt & Height Difference (Shim) Calculator

Tilt per 1 m = 0.04 mm/m · Height difference over L: Δh = 0.02 mm (20 µm) · Angle ≈ 8.3 arcsec

Formula: Δh (mm) = sensitivity s (mm/m) × divisions n × (L / 1000). With a 0.02 mm/m (≈ 4.1 arcsec) level over a 500 mm span, 1 division equals 0.01 mm (10 µm) of height difference—not 0.02 mm. The side toward which the bubble moves is the higher side.

Usage steps

  1. Checking the zero point (reversal method): place it on a reference surface plate and read the bubble, then reverse it 180° at the same spot and read again[2]
  2. If it reads the same position after reversal and is within the reference lines → no adjustment needed. Same position but outside the reference lines → the plate is tilted. Moves the same number of divisions left and right → correct the level's zero with the adjusting screw. Scattered → from the side with the larger error, correct half the maximum error on the level, then adjust the plate[2]
  3. When moving it on a workpiece, slide it rather than lifting it. Read only after the bubble has completely settled[2][4]
  4. How to measure straightness: place the level at the measuring position and read, slide it a little at a time and read through to the end, convert each reading to height and accumulate, and take the max minus min of the deviations from the straight line joining both ends as the straightness (11.2 µm in Obishi Keiki's example)[4]

Common measurement mistakes and countermeasures

Holding it by hand for a long time

WhyA high-sensitivity 0.02 mm/m level drifts in zero from the heat of the hand

CountermeasureDo not hold it in your hand continuously[2]

Storing it at extreme temperatures

WhyThe vial can break below −15 °C or above 40 °C

CountermeasureAvoid[2]

Care, storage, inspection and calibration

Care and storage

Inspection and calibration

Related instruments

Frequently asked questions about Precision level (spirit level)

What should I check first before measuring with Precision level (spirit level)?
Checking the zero point (reversal method): place it on a reference surface plate and read the bubble, then reverse it 180° at the same spot and read again Always wipe dust, burrs, and cutting oil from both the measuring faces and the workpiece before taking a reading.
What is the most common mistake when using Precision level (spirit level), and how do I prevent it?
Common error: Holding it by hand for a long time. A high-sensitivity 0.02 mm/m level drifts in zero from the heat of the hand Countermeasure: Do not hold it in your hand continuously
How should Precision level (spirit level) be cared for, stored, and inspected?
Care and storage: After use, rust-proof and put it in the storage box. Check sensitivity after a drop, etc. Inspection and calibration: Check accuracy periodically. Adjust the bubble size by tapping lightly with the secondary vial up (larger) or down (smaller)

Sources

Content without a mark was confirmed by two or more sources from different publishers. Procedures and cautions are summarized in Kezuriba's own words based on the sources, and the figures were drawn by Kezuriba. Accuracy values are manufacturers' published figures unless noted otherwise (values taken from JIS tables are marked as such).

  1. Japanese Standards Association (JSA), "JIS B 7510:1993 Precision levels (bibliographic record)"
  2. Obishi Keiki, "Precision Block Level (JIS B 7510 Grade A) Instruction Manual AA201"
  3. Obishi Keiki, "Precision Block Level (JIS B 7510 Grade A) Product Page"
  4. Obishi Keiki, "How to Measure Straightness with a Level"