Angles and flatness

Precision levelPrecision level (spirit level)

Also called: block level, flat-base level, level

Illustration of using a precision level

≡Standards, measuring range and accuracy

Standard
JIS B 7510:1993 Precision levels[1][2]
Measuring range
—Not confirmed in public sources
Resolution / graduation
  • Sensitivity 0.02 / 0.05 / 0.1 mm/m (grade A square type)
  • Definition The inclination needed to move the bubble one division. Expressed as height per 1 m of base. 1 arcsecond = 4.85 µm per m (≈ 5 µm). 0.02 mm/m ≈ 4 arcseconds, 0.05 mm/m ≈ 10 arcseconds, 0.1 mm/m ≈ 20 arcseconds
[2][3]1 source, for reference
Accuracy (instrument error / maximum permissible error)
—Not confirmed in public sourcesJIS tables not checked

▦Types

Square type (faces ensured parallel and square, with V-groove) / flat type

[2][3]1 source, for reference

◎Part names and functions

#NameEnglishFunction
1Main vial (bubble length adjustable with the air chamber)Main vial[2]
2Secondary vialCross vialChecks level in the roll direction when placed on a cylinder.[2]
3Adjusting screwAdjusting screwAdjusts the bubble position (zero point).[2]
4V-groove / base / gripV-groove, base, grip[2]

◷How to read the scale

  1. Read the divisions at both ends of the bubble and take the average[2][4]
  2. Get the height difference from the sensitivityExample: nominal 200 (length 200 mm), sensitivity 0.02 mm/m → 1 division = 4 µm/200 mm[4]

✋Usage steps

  1. Checking the zero point (reversal method): place it on a reference surface plate and read the bubble, then reverse it 180° at the same spot and read again[2]
  2. If it reads the same position after reversal and is within the reference lines → no adjustment needed. Same position but outside the reference lines → the plate is tilted. Moves the same number of divisions left and right → correct the level's zero with the adjusting screw. Scattered → from the side with the larger error, correct half the maximum error on the level, then adjust the plate[2]
  3. When moving it on a workpiece, slide it rather than lifting it. Read only after the bubble has completely settled[2][4]
  4. How to measure straightness: place the level at the measuring position and read, slide it a little at a time and read through to the end, convert each reading to height and accumulate, and take the max minus min of the deviations from the straight line joining both ends as the straightness (11.2 µm in Obishi Keiki's example)[4]

!Common measurement mistakes and countermeasures

Holding it by hand for a long time

WhyA high-sensitivity 0.02 mm/m level drifts in zero from the heat of the hand

CountermeasureDo not hold it in your hand continuously[2]

Storing it at extreme temperatures

WhyThe vial can break below −15 °C or above 40 °C

CountermeasureAvoid[2]

✓Care, storage, inspection and calibration

Care and storage

Inspection and calibration

The four patterns of the reversal method suit a diagram.

⇄Related instruments

📚Sources

Content without a mark is what agreed across two or more sources from different publishers. Procedures and cautions are summarized in Kezuriba's own words based on the sources, and the diagrams were drawn by Kezuriba. Accuracy values are manufacturers' published figures, not values from JIS tables.