Angles and flatness
Surface plateSurface plate
Also called: stone surface plate, granite surface plate, cast iron surface plate, Graplate (Mitutoyo product name)

Standards, measuring range and accuracy
- Standard
- JIS B 7513:1992 Precision surface plates
- ISO 8512-1:1990 (cast iron)
- ISO 8512-2:1990 (granite)
- Measuring range
- By manufacturer
- Obishi Keiki stone precision surface plate 300×300 mm to 1500×1000 mm and larger
- By manufacturer
- Accuracy (instrument error / maximum permissible error)
- Grade Grades 00, 0 and 1 (Obishi Keiki stone surface plates). ISO 8512 has grades 0, 1, 2, 3
- By manufacturer
- Obishi Keiki 300×300 mm Flatness: grade 00 2.0 µm, grade 0 4.0 µm, grade 1 8 µm
Metric ⇄ Inch & ASME/ISO Quick Reference: 1 mm = 0.03937 in (25.4 mm = 1 in) · 0.01 mm ≈ 0.0004 in (“4 tenths”) · 0.001 mm (1 µm) ≈ 0.00004 in (“40 millionths”). Standard reference temperature is 20 °C (68 °F) per ISO 1 / ASME B89.6.2. Related US standards: ASME B89.1.14 (calipers), ASME B89.1.13 (micrometers), ASME B89.1.10M (dial indicators), ASME B89.1.9 (gage blocks), ASME B89.3.7 (granite surface plates), ASME B1.2 (UN thread gages).
Types
Part names and functions
| # | Name | Function |
|---|---|---|
| 1 | Plate body (working surface) / support points (3) / auxiliary support points / adjusting bolts / lock nuts / stand | Supported at 3 points to level it.[4] |
Usage steps
- Installation: place it level on a stable stand with 3-point support (two points on the left, one on the right). Use the 2 auxiliary support points to prevent tipping[4]
- Leveling: loosen the lock nuts, retract the auxiliary support points so it rests on 3 points, place the level at the center of the left support point and then at the center of the plate, and level it with the adjusting bolts. Repeat this, tighten the lock nuts, and check again[4]
- For high-accuracy inspection, place it on a concrete base isolated from its surroundings, or on a sturdy stand on an isolated floor[4]
- Before use, clean the working surface and the workpiece and let them settle to the same temperature[4][7]
- Do not always use the same spot on the working surface; use the whole surface evenly[4]
Printed from Kezuriba (kezuriba.net/en/gauges/surface-plate/)
Common measurement mistakes and countermeasures
Placing it where vibration or temperature change is large
WhyIf a temperature difference develops between the top and bottom of the surface plate, it warps and its flatness changes; with vibration, instrument readings do not settle
CountermeasureAvoid[4]
Excessive load or impact
WhyLocal loads or impacts can chip the measuring surface or throw off its flatness
CountermeasureAvoid[4]
Care, storage, inspection and calibration
Care and storage
- Clean stone plates with glass cleaner, etc. (Mitutoyo introduces a PPC cleaner)[7]
- If it is scratched, send it for repair or inspection. If something has fallen on it, check for abnormalities[4]
Inspection and calibration
Related instruments
Frequently asked questions about Surface plate
- What should I check first before measuring with Surface plate?
- Installation: place it level on a stable stand with 3-point support (two points on the left, one on the right). Use the 2 auxiliary support points to prevent tipping Always wipe dust, burrs, and cutting oil from both the measuring faces and the workpiece before taking a reading.
- What is the most common mistake when using Surface plate, and how do I prevent it?
- Common error: Placing it where vibration or temperature change is large. If a temperature difference develops between the top and bottom of the surface plate, it warps and its flatness changes; with vibration, instrument readings do not settle Countermeasure: Avoid
- How should Surface plate be cared for, stored, and inspected?
- Care and storage: Clean stone plates with glass cleaner, etc. (Mitutoyo introduces a PPC cleaner) Inspection and calibration: Check accuracy (flatness) periodically. Obishi Keiki offers a calibration service
Sources
Content without a mark was confirmed by two or more sources from different publishers. Procedures and cautions are summarized in Kezuriba's own words based on the sources, and the figures were drawn by Kezuriba. Accuracy values are manufacturers' published figures unless noted otherwise (values taken from JIS tables are marked as such).
- Japanese Standards Association (JSA), "JIS B 7513:1992 Precision surface plates (bibliographic record)"
- ISO, "ISO 8512-1:1990 Surface plates — Part 1: Cast iron"
- ISO, "ISO 8512-2:1990 Surface plates — Part 2: Granite"
- Obishi Keiki, "Stone Precision Surface Plate Instruction Manual BA201"
- Mitutoyo, "Precision Granite Surface Plate (Graplate) Product Page"
- Obishi Keiki, "Stone Precision Surface Plate Product Page"
- Mitutoyo, "Check Points for Measuring Instruments (Catalog No. E12024)"
- Obishi Keiki, "How to Measure Straightness with a Level"