Surface and form
Coordinate measuring machine (CMM)Coordinate measuring machine (CMM)
Also called: coordinate measuring machine, CMM

Standards, measuring range and accuracy
- Standard
- JIS B 7440 series (Acceptance and reverification tests for coordinate measuring machines (CMM); -1:2003 Vocabulary, -2:2013 Length measurement, -5:2022 Probing, etc.)
- ISO 10360 series (JIS B 7440-2:2013 is ISO 10360-2:2009 adopted as a JIS without changing its technical content)
- Measuring range
- General From small benchtop models to large machines where the workpiece is placed on the floor. The measuring range is given for each model as the travel in X, Y and Z
- How to choose Choose the type by the size and weight of the workpiece and the surfaces you need to reach (stylus length)
- Accuracy (instrument error / maximum permissible error)
- Approach
- Length accuracy is expressed as the maximum permissible error of length measurement, E0,MPE. Length standards (gauge blocks, step gauges, etc.) are placed in 7 orientations within the measuring range (the 4 space diagonals are mandatory), and 5 lengths are each measured 3 times to verify it (JIS B 7440-2:2013)
- E0,MPE is often given in the form "A + L/K (µm, L is the length measured in mm)". Example: E0,MPE = 2.5 + 3L/1000 µm gives 4.0 µm at L = 500 mm (an example from Mitutoyo's explanatory material)
- In addition, there are the length measurement error E150,MPE with the stylus extended 150 mm sideways from the ram axis (added in ISO 10360-2:2009; in JIS B 7440-2 from the 2013 edition), R0, the width of scatter when the same length is measured 3 times, and the probing form error PFTU,MPE (a test sphere is measured at 25 points; the difference between the largest and smallest distances from the least-squares sphere; JIS B 7440-5)
- Pass or fail is decided taking the uncertainty of the test itself into account
- The MPE in a catalog is the value for a specified test. The uncertainty of an actual measurement changes with the arrangement of measuring points and how the datum is set, so a general estimate is difficult (Mitutoyo)
- Approach
Metric ⇄ Inch & ASME/ISO Quick Reference: 1 mm = 0.03937 in (25.4 mm = 1 in) · 0.01 mm ≈ 0.0004 in (“4 tenths”) · 0.001 mm (1 µm) ≈ 0.00004 in (“40 millionths”). Standard reference temperature is 20 °C (68 °F) per ISO 1 / ASME B89.6.2. Related US standards: ASME B89.1.14 (calipers), ASME B89.1.13 (micrometers), ASME B89.1.10M (dial indicators), ASME B89.1.9 (gage blocks), ASME B89.3.7 (granite surface plates), ASME B1.2 (UN thread gages).
Types
Moving bridge type
The X-axis carriage, Y-axis bridge and Z-axis ram move, and the workpiece sits on the base. Most models take this form (Mitutoyo)[3][4]
Fixed bridge type
The bridge is fixed and the table (Y axis) moves. Because it removes the error caused by a moving bridge, it is used on models aiming for higher accuracy[3]One source / reference
Horizontal arm type
A column, a vertically moving carriage and a horizontally extending ram. Some are high-speed models built into production lines, and some are space-saving models[3]One source / reference
Gantry (floor) type
Same construction as the moving-bridge type, with the workpiece placed on the floor. For large, heavy parts that do not fit on a table[3]One source / reference
Contact articulated-arm type
Points are taken with a probe at the end of a jointed arm[5]One source / reference
Part names and functions
| # | Name | Function |
|---|---|---|
| 1 | Stylus (tip ball) | The contact element attached to the end of the probe. The tip ball (ruby, silicon nitride, zirconia, etc.) is touched to the workpiece to get the coordinates of a point. High rigidity and a nearly perfect sphere are important[6][5] |
| 2 | Probe | The part that detects the stylus touching the workpiece and has the coordinates recorded at that moment. The stylus is attached with a screw (M2–M5, etc.)[6] |
| 3 | Master ball | A sphere with a certified diameter. Measuring it gives the center position and diameter of the stylus ball (calibration)[6][7] |
| 4 | X, Y, Z axes (carriage, bridge, ram) | Three axes that move the stylus to any position in space. On a moving-bridge model, the X axis is the carriage, the Y axis is the bridge and the Z axis is the ram. The positions of the three axes determine the coordinates of the touched point[3] |
| 5 | Guideways and air supply | The surfaces that serve as the reference for each axis's motion. On air-bearing models, the supply air pressure and the condition of the filter and mist separator also affect accuracy and life[7] |
| 6 | Software (PC) | Calculates features such as lines, circles, planes and cylinders from the points taken, and works out dimensions and geometric tolerances. The stylus registration values and coordinate systems are also managed here[5][3] |
How to read the scale
- Calculates features such as lines, circles, planes and cylinders from the point cloud, and finds dimensions and geometric tolerances such as flatness, parallelism, squareness and position[5]
- Even for the same workpiece and the same feature, the result changes with the number and arrangement of points and how the datum is set. When reading a result, also check which points it was calculated from[5][3]
Usage steps
- Equalize the temperature: dimensions are defined at 20 °C, so let the workpiece sit in the measuring room until it reaches room temperatureJIS B 7440-2 also has a section on environmental conditions for guaranteeing accuracy. In an example where a workpiece with a 15 °C temperature difference was placed in a temperature-controlled room, it settled in 80–100 minutes placed directly on a stone surface plate and in about 200 minutes held off the plate on 3 points (Oita Prefecture report). Placing it directly on a plate that carries heat away is faster[9][1][10]
- Start-of-day check: check the supply air pressure and clean the guideways and the master ball[7]
- Choose the stylus: as short as possible within reach, with as few extensions as possible. Choose a larger ball as long as the shaft does not hit the workpieceThe longer it is and the more extensions are added, the more accuracy drops from bending and deflection. A silicon nitride ball is recommended for aluminum, and zirconia for scanning cast iron[6]
- Calibration: measure the certified master ball and register the stylus ball's center position and diameter in the softwareAfter changing or reassembling the stylus, register it again before use. Only after registration are touched points corrected by the ball radius to give correct coordinates[6]
- Set the workpiece coordinate system: measure the drawing's datums (reference surfaces, lines, holes) and align the measurement origin and axis directions to the workpieceThe result changes with how the datum is set, so take it exactly as the drawing specifies[5][3]
- Measure: take points for each feature such as planes, circles and cylinders. For features where you want to check form (flatness, roundness, etc.), take more points and spread them over the surface or all the way around[5][3]
- Data processing: calculate distances, angles, parallelism, perpendicularity, position, etc. from the features and record them as the measurement result[5]
How errors come about
The "Common measurement mistakes" below cover what you end up doing wrong, while this section covers why the reading drifts and by how much.
1Machine-side errors, and errors the user controls
- A machine with 3 orthogonal axes has 21 geometric errors: 6 per axis (positioning, 2 straightness, pitch, yaw, roll) plus 3 squareness errors between axes. The maker measures these and compensates in software.
- What the user can control is temperature, the stylus, how points are taken, and workpiece fixturing. The NPL guide also lists a method of evaluating the scatter from fixturing by mounting and removing the workpiece 10 times.
- The error growing as the stylus extends sideways is the same idea as the Abbe principle. Bryan (1979) generalized it as: "Place the measuring system in line with the point to be measured. When that cannot be done, either make the guide free of angular motion or measure and correct the angular motion." Coordinate measuring machines and machine tools are designed and compensated on this idea.
2The measuring machine itself also expands with temperature
- The scale material and how it is mounted change the temperature effect. One study found that on a machine with a glass scale mounted apart from the body, the error changed almost in proportion to room temperature, while on a machine with a metal scale attached directly to the guide surface it hardly changed (Oya 1999).
- When the room temperature was changed from 20 °C to 25 °C, the air changed quickly, but it took more than 5 hours for the surface temperature of the measuring machine to settle (same source). Be careful with measurements on the day the air conditioning is switched, or after a door has been left open for a long time.
These are figures from sources with different conditions and are not meant to be compared on one scale. Use them only as a guide to the order of magnitude. The larger the temperature difference, and the larger the workpiece or machine, the longer it takes.
3Right after start-up, and lighting
- Even in a measuring room at 20±1 °C, coordinates drift by µm right after the machine is started or after the lights come on. LED lighting is a major cause of room temperature rise, and on some machines, when the machine and lights were turned on together in the morning, the coordinates did not settle to within 1 µm until evening (stable in about 1 hour with the lights off; Gunma Prefecture report).
- For measurements where 1 µm counts, turn on the machine and lights the day before.
[14]Reports from public institutions
Printed from Kezuriba (kezuriba.net/en/gauges/cmm/)
Common measurement mistakes and countermeasures
Measuring a workpiece right after machining or just brought in from another room
WhyBoth the workpiece and the machine expand and contract with temperature. Dimensions are defined at 20 °C
CountermeasureLet it reach room temperature in the measuring room before measuring. The effect of the temperature difference can be estimated with the thermal expansion calculator[9][1]
Not recalibrating after changing the stylus
WhyWhat is measured is the position of the stylus ball's center. If the ball's diameter and position are not registered correctly, the radius correction is off and shows up in the dimensions
CountermeasureRe-register with the master ball whenever you change the stylus. Keep the master ball clean[6]
Measuring a circle or plane with only 3 points
WhyWith 3 points, a circle or plane can always be fitted exactly, so form errors such as roundness and flatness do not show up in the numbers. The diameter and center values also change with the number and arrangement of points
CountermeasureFor features whose form you want to check, add more points and spread them around the whole circumference or surface. In the probing test (PFTU) as well, the test sphere is measured at 25 points[5][3]
Extending the stylus a lot with extensions, or sticking it far out sideways
WhyA long stylus bends easily. The farther it extends sideways from the ram axis, the more a slight tilt of the machine is magnified in the measured value (the same idea as the Abbe principle). E150 in JIS B 7440-2 is a test with the stylus extended 150 mm sideways. When it cannot be placed on one line, eliminate angular motion or measure and correct for it (Bryan's generalization)
CountermeasureKeep it as short as possible within reach and reduce the number of extension parts[6][3][11]
Moving fast with a long or heavy stylus
WhyDepending on the stylus length and weight and the machine's speed and acceleration, a point can be falsely triggered without the stylus touching the workpiece
CountermeasureReduce speed and acceleration with long or heavy styli[6]
Measuring in the Z direction (up and down) with a disk stylus
WhyA disk stylus is corrected for only one diameter, so it can only be used in the X and Y directions
CountermeasureFor the vertical direction, switch to another stylus (use a stylus changer)[6]
Treating the catalog MPE as the accuracy of a measurement result as is
WhyMPE is the value for a specified test. The actual uncertainty changes with the arrangement of measuring points and how the datum is set
CountermeasureRepeat important measurements with the same procedure and check the scatter. In addition, measure a calibrated workpiece that closely resembles the part with the same procedure, and derive the uncertainty of that measurement from the difference from the calibration value and the scatter (ISO 15530-3)[3][12][13]
Measuring right after start-up or right after turning the lights on
WhyEven in a room at 20±1 °C, the machine temperature changes right after start-up or after the lights come on, and coordinates drift by µm. In one case it took almost half a day to settle
CountermeasureFor measurements where 1 µm counts, turn on the machine and lights the day before. For the first measurement in the morning, check with a standard on hand first[14]
Care, storage, inspection and calibration
Care and storage
- Check the supply air pressure (at start of work)[7]
- Replace the air filter and mist separator elements periodically[7]
- Clean the guideways. Keep the master ball clean too[7]
- Check the stylus ball for wear. Ruby balls can wear when scanning aluminum or cast iron, and silicon nitride wears heavily on steel surfaces[6]
Inspection and calibration
- Length accuracy: per JIS B 7440-2 (ISO 10360-2), check E0 (and E150 if needed) and the repeatability R0 with length standards. Use the same method for acceptance tests and reverification tests[1][3]Standard
- Probing: per JIS B 7440-5, measure a test sphere at 25 points to check PFTU. Before the length test, it is desirable to confirm that probing works to specification (foreword of JIS B 7440-2)[1][3]Standard
- Between periodic inspections, do interim checks with a standard on hand (JIS B 7440-2 Annex A, informative). Example: periodically measure a ball dimension gauge of low-thermal-expansion ceramic or carbon tool steel with the same placement and the same program, and watch for changes in the values (Wago et al. 2013)[1][15]StandardPaper
- When you have calibration done externally, ask to have the measurement results kept both as received (before adjustment) and after adjustment. Also decide in the contract which test items are performed (Mitutoyo America explanatory material)[8]
A CMM is a measuring machine that, unlike a caliper or micrometer, does not read a dimension at one place but calculates form and position from the coordinates of points. The symbols and meanings of geometric tolerances can be looked up on the geometric tolerance page under "Drafting".
Related instruments
Frequently asked questions about Coordinate measuring machine (CMM)
- What should I check first before measuring with Coordinate measuring machine (CMM)?
- Equalize the temperature: dimensions are defined at 20 °C, so let the workpiece sit in the measuring room until it reaches room temperature JIS B 7440-2 also has a section on environmental conditions for guaranteeing accuracy. In an example where a workpiece with a 15 °C temperature difference was placed in a temperature-controlled room, it settled in 80–100 minutes placed directly on a stone surface plate and in about 200 minutes held off the plate on 3 points (Oita Prefecture report). Placing it directly on a plate that carries heat away is faster Always wipe dust, burrs, and cutting oil from both the measuring faces and the workpiece before taking a reading.
- What is the most common mistake when using Coordinate measuring machine (CMM), and how do I prevent it?
- Common error: Measuring a workpiece right after machining or just brought in from another room. Both the workpiece and the machine expand and contract with temperature. Dimensions are defined at 20 °C Countermeasure: Let it reach room temperature in the measuring room before measuring. The effect of the temperature difference can be estimated with the thermal expansion calculator
- How should Coordinate measuring machine (CMM) be cared for, stored, and inspected?
- Care and storage: Check the supply air pressure (at start of work) Inspection and calibration: Length accuracy: per JIS B 7440-2 (ISO 10360-2), check E0 (and E150 if needed) and the repeatability R0 with length standards. Use the same method for acceptance tests and reverification tests
Sources
Content without a mark was confirmed by two or more sources from different publishers. Procedures and cautions are summarized in Kezuriba's own words based on the sources, and the figures were drawn by Kezuriba. Accuracy values are manufacturers' published figures unless noted otherwise (values taken from JIS tables are marked as such).
- Japanese Standards Association (JSA), "JIS B 7440-2:2013 preview (table of contents and foreword; identical to ISO 10360-2:2009)"
- Japanese Standards Association (JSA), "JIS B 7440 series: Acceptance and reverification tests for coordinate measuring machines (bibliographic record)"
- Mitutoyo, "Fundamentals of Coordinate Measuring Machines"
- Mitutoyo, "Quick Guide to Precision Measuring Instruments (Catalog No. E11003(7), as of August 2024)"
- Saitama University, "Machine Shop Practice: Coordinate Measuring Machine Textbook"
- Mitutoyo, "Stylus basics for CMMs (selection, materials, calibration)"
- Mitutoyo, "CMM maintenance video (supply air pressure, air filter, guideways, master ball)"
- Mitutoyo, "Mitutoyo America 'CMM Calibration and Performance Testing' explanatory material (2016, E0, E150, R0 and PFTU of ISO 10360)"
- ISO, "ISO 1:2022 GPS — Standard reference temperature (standard reference temperature for dimensions: 20 °C)"
- Oita Prefectural Industrial Research Institute, "Goto, Motoyama, Otsuka, 'Study of measurement errors due to the temperature environment in coordinate measuring machines,' FY1986 research report (1986)"
- P. Schellekens et al. (Annals of the CIRP 47(2), 1998), "Design for Precision: Current Status and Trends"
- ISO (iTeh preview), "ISO 15530-3:2011 Use of calibrated workpieces or measurement standards"
- National Physical Laboratory (NPL, UK), "D. Flack, 'Good Practice Guide No. 130: CMM task-specific measurement uncertainties'"
- Gunma Industrial Technology Center, "Kano et al., 'Study on the effect of ambient temperature on coordinate measurement accuracy,' research report (2019)"
- Wago et al. (Journal of the Japan Society for Precision Engineering 79(3), 2013), "A study of interim check methods for coordinate measuring machines"
- H. Schwenke et al. (CIRP Annals 57(2), 2008), "Geometric error measurement and compensation of machines — An update (abstract)"
- Applied Sciences 14(7) 2982 (MDPI, 2024), "Efficient Method for Identifying Key Errors Based on 21-Geometric-Error Measurement of Three Linear Axes of Machine Tools"
- Oya (Transactions of the JSME, Series C, 65(631), 1999), "Effect of the temperature environment on the accuracy of coordinate measuring machines (changes in error distribution in environments with different room temperatures, and their compensation)"