Measuring by comparison

Lever-type dial indicator (dial test indicator)Dial test indicator (lever type)

Also called: test indicator, lever test, pick test (colloquial)

Illustration of using a lever-type dial test indicator

Standards, measuring range and accuracy

Standard
  • JIS B 7533:2015 Lever-type dial gauges
  • ISO 9493:2010 Dial test indicators (lever type)
[1][2][3][4]
Measuring range
  • By manufacturer
    • Mitutoyo TI-113HX (513-404-10H) 0.8 mm (graduation 0.01 mm)
[6][9]One source / reference
Resolution / graduation
  • General 0.01 / 0.002 / 0.001 mm (metric)
  • Guide to stylus ball diameter 1 mm: very small features, 2 mm: general use, 3 mm: when you want stable readings on a rough surface
[5][3]One source / reference
Accuracy (instrument error / maximum permissible error)
  • Approach JIS B 7533:2015 specifies the forward-direction indication error over the full measuring range, the 10-division indication error, the 1-revolution indication error, return error, repeatability, and measuring force. The forward direction is the direction against the measuring force
  • By manufacturer
    • Mitutoyo 513-404-10H (0.01 mm, 0.8 mm) 10 divisions 5 µm, full range 9 µm, return error 4 µm, repeatability 3 µm, measuring force 0.3 N or less
[3][6]One source / reference

Metric ⇄ Inch & ASME/ISO Quick Reference: 1 mm = 0.03937 in (25.4 mm = 1 in) · 0.01 mm ≈ 0.0004 in (“4 tenths”) · 0.001 mm (1 µm) ≈ 0.00004 in (“40 millionths”). Standard reference temperature is 20 °C (68 °F) per ISO 1 / ASME B89.6.2. Related US standards: ASME B89.1.14 (calipers), ASME B89.1.13 (micrometers), ASME B89.1.10M (dial indicators), ASME B89.1.9 (gage blocks), ASME B89.3.7 (granite surface plates), ASME B1.2 (UN thread gages).

Types

Horizontal, vertical, inclined, parallel, universal (swiveling contact point), pocket, and long-stylus types

[5][6]One source / reference

Types where a switching lever changes the measuring direction

[7][4][8]

Part names and functions

#NameFunction
1Contact point (stylus)Moves as part of the lever, with the ball at its tip touching the workpiece. The angle can be changed over a range of about 200°.[9][4][10][8]
2Pointer and dial faceThe scale is mostly symmetrical about 0 (balanced scale).[4][8][10]
3Reversing lever (direction switch)Switches the measuring direction (push/pull).[4][8]
4Dovetail plate (dovetail) and stemThe mounting part for a height gauge or stand.[4][8][5]
5Case and cover glassTurn the outer bezel to align the dial's 0 with the pointer. A transparent cover protects the dial and pointer[4][8]

How to read the scale

  1. Cosine correction: when the angle θ between the stylus and the measuring direction is large, multiply the reading by a correction factor (cos θ)Example: reading 0.002 mm, θ = 30° → 0.002 × 0.87 = 0.00174 mm (example from Mitutoyo E11003)[3][10]
  2. Correction factor table (Mitutoyo): 10° 0.98, 20° 0.94, 30° 0.86–0.87, 40° 0.76, 50° 0.64, 60° 0.50[10][3]
  3. Comparison reading from the zero point with graduation 0.01 mm and measuring range 0.8 mm (Mitutoyo's explanation has an example diagram)[9]
Cosine error of a lever-type dial test indicator: angle θ between the stylus and the workpiece surface
Angle θ between the contact point and the surfaceTrue displacement = reading × cos θAt θ = 30° the true value is 0.87 times the readingBelow 10°, no correction is considered necessaryθMeasuring direction (the surface moves up and down)Part

If the contact point is parallel to the workpiece surface (perpendicular to the measuring direction), the reading is used as is. With an angle θ the reading comes out larger, so correct the true displacement with "reading × cos θ" (calculation below). The picture is a conceptual illustration generated in Canva.

Cosine correction calculation

Correction factor cos θ = 0.866 Corrected value = 0.00173 mm

This is the calculated value of cos θ. Manufacturers' factor tables use either 0.86 or 0.87 for 30° (cos 30° = 0.866). If the angle is smaller than 10° (15° in some sources), correction is said to be unnecessary.

Usage steps

  1. Operation check: lightly move the stylus with a finger and check that the pointer follows smoothly and the return is not slow. Check that the stylus is securely fixed to the bearing[9][4]
  2. Check that the stem or dovetail stem is not loose (TECLOCK gives the tightening torque of the contact point screw as 1.5–2 kg·cm (about 15–20 N·cm))[9][4]
  3. Mounting: secure it in a rigid holder, square to the front so it does not tilt[9][10][5]
  4. Contact point angle: set the contact point at a right angle to the measuring direction (the direction of displacement), i.e. parallel to the measured surface[9][4][8][3]
  5. Zero point: setting zero at the middle of the measuring range lets you read unexpected dimensions within range[9]
  6. Uses: runout of rotating parts, parallelism against a precision surface plate, comparison with a reference, centering on a lathe, setup on a milling machine, etc.[9][5]

Using it for centering

Step-by-step for four-jaw chucks, bores and bosses, and tailstocks, with a calculator that finds how far to move from four readings →

Common measurement mistakes and countermeasures

Cosine error (contact point angle)

WhyWith a lever-type indicator, the magnification changes with the angle between the contact point and the direction the workpiece moves; it reads 1:1 only when they are aligned

CountermeasureKeep the angle as small as possible. Mitutoyo E12024 says it is negligible below 10°, and Mitutoyo's US material recommends below 15°. At larger angles, multiply by a correction factor[10][3][9][4][8]

Using a contact point of a different length

WhyThe lever ratio changes, causing a large error

CountermeasureUse the contact point length specified for each model. Long contact points are easy to knock, so use them only when needed[10][9][4][5]

Force from directions other than the measuring direction

WhyCauses performance degradation or damage

CountermeasureDo not apply side force[4]

Care, storage, inspection and calibration

Care and storage

Inspection and calibration

For the correction factor at 30°, Mitutoyo's own documents give both 0.86 (E12024) and 0.87 (E11003). The difference is just how cos 30° ≈ 0.866 is rounded.

Related instruments

Frequently asked questions about Dial test indicator (lever type)

What should I check first before measuring with Dial test indicator (lever type)?
Operation check: lightly move the stylus with a finger and check that the pointer follows smoothly and the return is not slow. Check that the stylus is securely fixed to the bearing Always wipe dust, burrs, and cutting oil from both the measuring faces and the workpiece before taking a reading.
What is the most common mistake when using Dial test indicator (lever type), and how do I prevent it?
Common error: Cosine error (contact point angle). With a lever-type indicator, the magnification changes with the angle between the contact point and the direction the workpiece moves; it reads 1:1 only when they are aligned Countermeasure: Keep the angle as small as possible. Mitutoyo E12024 says it is negligible below 10°, and Mitutoyo's US material recommends below 15°. At larger angles, multiply by a correction factor
How should Dial test indicator (lever type) be cared for, stored, and inspected?
Care and storage: Wipe dirt off the contact point with a dry cloth or a cloth with a little alcohol. Do not oil it Inspection and calibration: JIS B 7533 method: hold it, set zero near the starting point, read the indication error every 10 divisions in the forward direction, move past the end point, then read in the return direction as well

Sources

Content without a mark was confirmed by two or more sources from different publishers. Procedures and cautions are summarized in Kezuriba's own words based on the sources, and the figures were drawn by Kezuriba. Accuracy values are manufacturers' published figures unless noted otherwise (values taken from JIS tables are marked as such).

  1. Japanese Standards Association (JSA), "JIS B 7533:2015 Lever-type dial gauges (bibliographic record)"
  2. ISO, "ISO 9493:2010 Dial test indicators (lever type)"
  3. Mitutoyo, "Quick Guide to Precision Measuring Instruments (Catalog No. E11003(7), as of August 2024)"
  4. Teclock, "General Catalog 1: Dial Indicators (cautions for use, including JIS B 7503 part drawings)"
  5. Mitutoyo, "Mitutoyo Product Fundamentals (Bulletin No. 2180, Mitutoyo America)"
  6. Mitutoyo, "Test Indicator TI-113HX 513-404-10H Specifications"
  7. Teclock, "General Catalog 2: Lever Tests"
  8. Niigata University, "Niigata University Faculty of Engineering, Technical Staff Training Text, Chapter 1: Workshop Measurement"
  9. Mitutoyo, "How to Use Test Indicators (Lever-Type Dial Gauges) Correctly and Cautions"
  10. Mitutoyo, "Check Points for Measuring Instruments (Catalog No. E12024)"
  11. Oita Prefectural Industrial Research Institute, "On Maintaining the Accuracy of Dial Indicators Used by Companies in the Prefecture (1991)"